AgentSlop ResearchBIOSCIENCE  /  FIELD REPORTSDevelopment study

Perinatal signal studies / 01

CTG Outcome Atlas

Can the final 30 minutes of a fetal heart-rate and uterine-activity recording flag umbilical arterial pH below 7.10?

Retrospective only. Selected after many comparisons on 441 development recordings. Not independently validated.
FIG. 01   Cohort atlasView result #470

What did the model flag?

Selected example · XGBoost #15 · 30-minute input

One mark per recording; grouped by outcome, not time

45 with low pH

Detected
33
Missed
12

396 without low pH

False alerts
76
Not flagged
320
33 of 45

low-pH recordings detected 73.3% sensitivity

33 of 109

alerts were true positives 30.3% of alerts

Measured FPR: 76 false alerts / 396 without low pH = 19.19%.

The price of detection.

What changes when more false alerts are allowed? Compare five representative result rows from the same 441 development recordings. Sensitivity is measured among 45 low-pH recordings; the false-positive rate (FPR) among the other 396.

Two ways to read the same 33.

Selected 30-minute row · same detections, different denominators.

Among 45 low-pH recordings

73.3%

33 detected · 12 missed

Among 109 model alerts

30.3%

33 true · 76 false alerts

Exploratory, not independently validated. The same development split was reused for many runs; selecting a winner here can overstate performance on new data. Why selection on cross-validation results can bias estimates. TRAIN20 is a training-side target, not a promise about measured development FPR. At the same sensitivity, fewer false positives break ties. A derived result is not an independent fit.

Table · operating points at all three caps

One 30-minute input vs. sixteen 15-minute views

Observed FPR cap30-minute inputSixteen 15-minute views
<20% or ≤20%#15: 33/45 · 76/396 FP#13 and #15: 27/45 · 79/396 FP
<25%#132: 34/45 · 92/396 FP#57: 31/45 · 86/396 FP (matched by derived #59)
≤25%#30: 35/45 · 99/396 FP#57: 31/45 · 86/396 FP (matched by derived #59)

The leaders differ by input. This does not establish a causal effect of shortening the input; compare matched pairs for that question. The short-window winners at 20% used fixed views; the CNN at 25% trained on random crops and was scored on fixed views. Sixteen views together cover the same 30 minutes, so this is not a 15-minute lead-time result.

Compare input representations

Same evaluation cohort, different inputs and training

One trace or sixteen views of the same half hour?

Both arms were evaluated on the same 441 development recordings. Sixteen overlapping 15-minute views, starting one minute apart, still cover the full half hour. Differences in fitting, training exposure and threshold calibration mean these are not controlled estimates of the effect of window length. The deltas compare aggregate counts, not which cases changed.

30 minutes: 33/45 detected, 76/396 false alerts. Sixteen overlapping views: 27/45 detected, 79/396 false alerts. Enable JavaScript to compare the training and threshold notes for all four pairs.

Table · ranking and probability-error metrics

Ranking and probability error have different leaders

MetricBest recorded valueExperiment
AUROC ↑0.8380SoftTail10-R1 · row 534
Average precision ↑0.4034#132 SERIAL4-Q · row 38
Brier score ↓0.0757#166 I2 · row 522
Log loss ↓0.2572#172 T saved30 · row 499

Values use the precision of the embedded source rows. Metric leaders are examples; tied values and shared variants can appear elsewhere in the tracker. These are reported full-count results from mixed protocols, not a head-to-head or a model selection recommendation.

One recording, one outcome.

Cardiotocography (CTG) records fetal heart rate and uterine activity during labour. The question is whether these signals can flag umbilical arterial cord-blood pH below 7.10 at birth. Exactly 7.10 counts as negative. The study does not diagnose hypoxia or brain injury.

FIG. 02   /   Synthetic signal

A moving 15-minute window.

Illustration only.
Not a patient recording.

FHR bpmTOCO rel.
Illustrative fetal heart rate and uterine activity over thirty minutesA synthetic fetal heart rate trace in beats per minute above a slower relative uterine-activity trace. These curves are an instrument illustration, not the inputs for a particular prediction.
051015202530 min
FHR: fetal heart rateTOCO: uterine activity (relative units)
View 1 / 16 · minutes 0–15
How the sixteen windows overlap

01 / Signal window

Two views of the same half hour

Both inputs draw from the final 30 minutes of recorded fetal heart rate and uterine activity.

The first short window covers minutes 0–15, the next 1–16, and the last 15–30. Together they span 30 minutes, not 16 independent recordings. The signal end is not necessarily birth.

02 / Validation

Which recordings were scored?

One hospital in Brno contributed 552 selected recordings to CTU-UHB. Recordings were split before making windows, so all views of a recording stayed together.

552 source recordings Selected at one centre
441 Development
45 low-pH / 396 other. Five outer folds, nine low-pH recordings in each.
111 Locked
Reserved and unscored. These recordings come from the same centre.

Bar lengths are proportional to cohort size. The reserved set is not external validation.

The registry repeatedly compares methods on those same 441 recordings. One extra event caught changes sensitivity by 2.22 percentage points.

Threshold example · why ten extra detections cost 62 more false alerts

03 / Threshold example · #44

What did ten extra detections cost?

The same 30-minute SmallCNN scores were read at two cutoffs. TRAIN20 targets 20% false positives in training; it does not guarantee 20% on other recordings.

SmallCNN development outcomes at two decision thresholds
CutoffLow pH detectedFalse positives
Default 0.525/45 55.6%54/396 13.6%
TRAIN2035/45 77.8%116/396 29.3%

10 more events caught, 62 more false alerts. Ranking did not change: AUROC was 0.7995 at both cutoffs.

Which IDs have full-development results?

192/192 registry IDs have at least four measured rows at their reported scope. 191/192 have four across the full 441-record development set. The exception, #46, covers 177 recordings (18 events), not all five folds.

An ID can be a model, a decision rule, a replica or an alias. These counts do not represent 192 independent models or replications.

What would move this beyond a retrospective benchmark?
  1. Freeze the model and cutoff before a single evaluation on the locked 111. A small reserved cohort limits the precision of sensitivity estimates.
  2. Test at another centre with a defined time before delivery and report both missed events and false alerts.
  3. Test the intended input and hardware. A waveform ONNX export is not evidence of accurate camera digitisation or offline phone performance. Those need separate tests on actual photos and devices.

These are next steps. No clinical deployment, intervention benefit or individual risk estimate is established.

See the source dataset and the versioned aggregate results.

Same weights, different alerts.

Two entries share 20 fetal heart-rate CNN fits. #57 reads the pooled scores directly; #59 corrects those same scores. The experiment changes the readout, not the network.

At the fixed reference cutoff

185 → 0alerts, raw → corrected

At training-selected cutoffs

117 → 117alerts, raw → corrected
FIG. 03   Score lineage Sixteen 15-minute views · 441 held-fold predictions per row

Read the last column: both cutoffs selected on training data produced 31 detections out of 45 and 86 false alerts out of 396 (21.7% observed FPR). The matching totals are not two independent confirmations. Lines show shared model weights, not causal effects.

Calibrating the CNN narrowed the gap

Brier score measures probability error; lower is better. These are development comparisons.

.13.10.07Raw probabilitiesNested calibration
CNN, solid
0.12484 → 0.08206
Logistic regression, dashed
0.08060 → 0.08021

The CNN’s probability error fell, but did not beat the calibrated logistic baseline.

Earlier baselines, language models and imaging experiments

Initial ranking baselines

  • Waveform CNN: 0.799 AUROC, 0.291 average precision.
  • Logistic regression: 0.792 AUROC, 0.354 average precision.
  • XGBoost: 0.793 AUROC, 0.289 average precision.

Language-model and imaging experiments

  • Frozen Decider #24: default threshold raised no alerts. Its training-selected threshold caught 14/45 on 30 minutes and 19/45 with sixteen views.
  • Kev #42: its fixed 0.5 threshold detected no events in either arm. In the original matched-feature test, adapted Kev had a higher (worse) Brier score than calibrated logistic regression: 0.09158 vs. 0.08021.
  • Rendered plot readability is not diagnosis, extraction accuracy, or real phone-photo validation.

Next evidence needed: prespecify a model and threshold, then test on genuinely independent data. Offline deployment, real-image reliability, calibration, and clinical utility have not been established. Independence at the mother level was not established.

Evidence workbench.

Sort the archive by any result column. Open a result number for its inputs, model, training notes, and statistics.

192registered IDs
1,194registry-linked rows
27supplementary rows

Query the register

Aggregate evidence only

191 IDs have full-development coverage. Compare rows within the same cohort and protocol. Sensitivity and false-positive rate sort by percentage. Export includes all matching results.

Page 1

From waveform to phone.

Interface concept · synthetic signals

Planned device research

When should the app decline to score?

Start with the same fetal heart-rate and uterine-activity chart. Check whether both channels and the required time span can be recovered reliably.

Quality check fails
Return “Unable to score”. Do not display a risk estimate.
Quality check passes
In a future study, compare Decider with waveform ML on the same eligible inputs.
See what has been tested ↓
Evidence boundary

One verified component, not a verified app.

Only the waveform model export has a parity check. The combined photo-to-score workflow and a head-to-head scoring test are still planned.

  1. 01 / Capture

    Photograph a chart

    Test actual paper traces under glare, tilt and imperfect framing.

    Real photos untested
  2. 02 / Extract

    Recover and check both signals

    Check usable coverage, time alignment and extraction quality. If a required channel or segment fails, return “Unable to score.” Define and validate the quality criteria before testing.

    End-to-end untested
  3. 03 / Score

    Compare two scoring approaches

    Planned: Decider vs. waveform ML on the same extracted signals and eligible recordings. Report missed events, false alerts and how often scoring is declined.

    Waveform ONNX parity only · #30
  4. 04 / Deploy

    Measure on a phone

    Test offline operation, latency and memory use on physical hardware.

    Device untested

Readability is a separate result. Decider appears here as a planned comparison only, not as an app result or a new training run. Tests on waveform-rendered plots (#29) with simulated distortions do not establish real-photo extraction accuracy, on-device performance or clinical utility.

Sources and boundaries.

How to read the result rows
  • TP and FN partition the event count; FP and TN partition non-events. Sensitivity = TP ÷ (TP + FN); FPR = FP ÷ (FP + TN).
  • AUROC and average precision summarize ranking. Brier score and log loss evaluate probabilities; lower is better. A missing metric is not zero.
  • Reported cohort sizes and threshold policies vary. Select 441 for full-development rows and compare like with like. Tracker row numbers are provenance, not patient IDs.